small logo

ISE: Carbon Dioxide

ISE: Carbon Dioxide
(Model # CO21501)

 
6103 GLENMONT · HOUSTON, TEXAS · 77081
800/522-7920 · FAX: 713/772-4671 · 713/772-6641
www.VL-PC.com
 
Carbon Dioxide Electrode Model CO21501
SLOPE55 + / - 3 mV/decade
REPRODUCIBILITY+ / - 2%
INTERFERENCESVolatile weak acids
COMPLEXATIONHeavy metals
TEMPERATURE RANGE0 to 50° C (if calibrated at that temperature)
PRESSURE RANGE0 to 10 psi
RESPONSE TIME95% response in 30 seconds
STORAGELong term, store dry
Short term, store in carbon dioxide standard with ISA added
CONCENTRATION RANGE10-4 to 10-2M
 
4.4 ppm to 440 ppm
pH RANGE4.8 to 5.2 pH
RESISTANCE1000 to 1500 megaohms
CARE, MAINTENANCE, & CLEANINGTreat with dilute carbon dioxide standard with pH adjusted to 4.8 - 5.2
Replace membrane as needed
TEMPERATURE COMPENSATIONNo
ISOPOTENTIAL POINTUnknown
ON-LINE CAPABILITYYes, pH control required
APPLICATIONSSoft Drinks, Wine Fermentation

FOR ADDITIONAL PRODUCT INFORMATION,
CALL 713/772-6641 OR 800/522-7920
Email: sales@VL-PC.com
www.VL-PC.com


 
©2008 Van London-pHoenix Company