The glass body Dissolved Oxygen electrodes are available for laboratory fermentation, where sterilization takes place in a separate autoclave. The disposable Van London - Phoenix Co. Dissolved Oxygen Electrodes are compatible with most common oxygen amplifiers.
|
 |
This series of steam sterilizable, Dissolved Oxygen electrodes are designed for economical use. They are constructed very similarly to the standard pH electrode, and have a 10 and 12mm body, and can be ordered with the desired “H” length. The “H” length is determined by the insertion depth on the under side of the fermentor headplate to the tip of the probe. Glass disposable Dissolved Oxygen electrodes are available in both polarographic and galvanic styles to fit all requirements.
The glass body Dissolved Oxygen probes are available for laboratory fermentation, where sterilization takes place in a separate autoclave. The disposable Van London - Phoenix Co. Dissolved Oxygen Electrodes are compatible with most common oxygen amplifiers. Glass body Dissolved Oxygen electrodes are non-temperature compensated and are required to be kept at a constant temperature. Calibration must be made at the same temperature as the controlled fermentation. The electrodes feature external filling tubes. Electrolyte can be added during use without removing the electrode from the vessel. The electrode membranes can be replaced very easily. However, some skill is involved in replacing them.
|
| |
|
| Type |
027IP15
Polarographic |
027BG15
Galvanic |
027NG15
Galvanic |
| CATHODE |
Pt |
Ag |
Ag |
| ANODE |
Ag/AgCl |
Pb |
Pb |
| RESPONSE |
98% in 60 seconds |
98% in 90 seconds |
98% in 90 seconds |
| STABILITY* |
Better than 2% |
Better than 2% |
Better than 2% |
| TEMPERATURE COMPENSATION |
NO |
NO |
NO |
| TEMPERATURE RANGE |
0-60°C |
0-60°C |
0-60°C |
| STERILIZATION TEMPERATURE |
130°C |
130°C |
130°C |
| RESIDUAL SIGNAL |
Less than 1 nA |
0.5 - 1 mV |
Less than 0.4 µA |
| OUTPUT** |
50 nA |
40 mV ± 5 mV |
10-15 µA |
| |
| * = Constant Temperature |
** = Medium Saturated With Air |
|
| |
| ORDERING INFORMATION |
| |
| PART NO. |
"H" LENGTH |
BODY DIAMETER |
TYPE |
| 027IP15 |
"H"-DL |
12 mm |
Polarographic |
| *027BG15 |
"H"-DL |
12 mm |
Galvanic |
| 027IP15 |
"H"-DL |
10 mm |
Galvanic |
|
| |
| * Must be used with 9001094-010U. |
| |
| MEMBRANE SERVICE KITS AND FILLING SOLUTION |
| |
| PART NUMBER |
DESCRIPTION |
| M0500T1-K5 |
Membrane Service Kit K-5 for both Polarographic and Galvanic O2 electrodes. Consists of 10 membranes 1 mil thick and 10 silicone bands |
| M0500T2-K5 |
Membrane Service Kit K-5 for both Polarographic and Galvanic O2 electrodes. Consists of 10 membranes 2 mil thick and 10 silicone bands |
| R001068 |
Polarographic O2 electrode filling solution, 125 mL. |
| R001069 |
Galvanic O2 electrode filling solution, 125 mL. |
| |
| CABLES |
| For non-temperature compensated O2 electrodes, used with Polarographic and Galvanic O2 electrodes. Standard 3 meter lead. |
| |
| PART NUMBER |
DESCRIPTION |
| 9001091-010B |
BNC Connector |
| 9001091-010I |
DIN Connector |
| 9001091-010D |
Dual Pin Connector |
| 9001091-010U |
Banana Plugs |
| 9001091-010T |
No Connector |
| 9001091-010CC |
Custom cable or for use when customer supplies connector |
| 9001093-010D |
Pin Connector. Cable features a load resistor for direct connection to a recorder with mV output. (Approximately 40 mV saturated air/water) |
|
|
|
|