Refillable pH and ORP Electrodes TEXT SIZE: A A A    PRINT THIS PAGE   EMAIL THIS PAGE

This series of steam-sterilizable pH and ORP electrodes are designed for small fermentors where sterilization is performed in an autoclave in non-pressurized operation. The electrodes can withstand repeated sterilization up to 130°C and feature a detachable cable assembly insulated with a water tight connector.

Steam Sterilizable pH and ORP electrode

This series of steam-sterilizable pH and ORP electrodes are designed for small fermentors where strerilization is performed in an autoclave in non-pressurized operation. The electrodes can withstand repeated sterilization up to 130°C and feature a detachable cable assembly insulated with a water tight connector. The electrodes are autoclaved with the cable detached and a protective cap over the connector. The electrode construction features an annular ceramic reference junction to eliminate junction problems which normally occur with autoclavable steam-sterilizable electrodes. Electrodes are available in both single and double junction reference chambers. The double junction reference style is recommended in most cases for long term stability and life expectancy. Both reference chambers are refillable with a special high temperature reference electrolyte. When ordering, the detachable cable must be ordered separately with the proper pH meter connection. Unless otherwise specified, the immersion length is 150mm.
The immersion length “A” is determined by the immersion depth “H” of the probe plus 70mm.

Features
1. Removable electrode cable connection.
2. Internal Ag/AgCl reference chamber fill port.
3. Electrolyte bridge fill port.
4. Internal annular ceramic reference junction.
5. Annular ceramic primary reference junction.
6. pH bulb or platinum band sensor.
7. 12mm O.D. shaft.

Electrodes
A774713-XXX-DL Double reference chamber, GXV pH bulb, 0-14 pH
A779861-XXX-DL ORP Double reference chamber
 
Detachable Leads and Accessories
9001081-XXX-X 5mm O.D. low-noise DL cable R001026 Inner reference fill 3M KCl saturated with AgCl. 125 ml
9001083-XXX-X Miniature 3mm O.D. low-noise DL cable R001027 Outer reference fill 3M KCl. 125 ml
9001082-XXX-X Low-noise triax DL cable R001028 Outer reference fill 1M KNO3. 125 ml

 

Refillable In-Situ pH and ORP Electrodes
Autoclavable Pressurized Application
For Sterilizable, Pressurized Fermentors with Non-Removable Electrode

 

In-Situ pH and ORP electrode

In-Situ pH and ORP Electrodes

This series of pH and ORP electrodes are designed for use in pressurized immersion electrode assemblies used in fermentors and reaction vessels. The sterilizable combination electrodes are designed for repeated steam sterilization cycles at 130°C. Pressure compensation should be effected during all sterilization cycles. The electrodes are constructed with an annular ceramic junction eliminating the fouling problem which normally occurs with sterilizable electrodes. Electrodes are available in both single and double junction chambers. The double junction electrode is recommended over the single junction for long-term stability and use. By the virtue of this special design, the internal reference junction is a sealed Ag/AgCl reference with a special high temperature KCl gel reference solution. An internal annular ceramic junction eliminates fouling problems. The outer junction is refillable and the electrolyte may be replaced and changed during use. Both the single and double junction styles require a special high temperature gelled reference electrolyte solution. The DL series (detachable lead) electrode is used in the VLC probe assemblies P/N 4051001. Order the cable separately.

Unless otherwise specified, the immersion length “A” is 120mm. The immersion length “A” is determined by the immersion depth “H” of the probe, plus 50mm.

Features
1. Standard fixed cable series.
2. Optional removable lead, DL series.
3. Sealed gel-filled Ag/AgCl internal reference cell.
4. Gel-filled outer reference electrolyte.
5. Internal annular ceramic reference junction.
    Double junction only.
6. Annular ceramic primary reference junction.
7. pH bulb or platinum sensor.

 

Electrodes
B773792-____-X39   Double reference chamber, GXV pH bulb, -5° to +135°C, also available with DL
B779861-____-X39   ORP double reference chamber, -5° to +135°C, also available with DL

 

When ordering, determine the immersion length "A" in millimeters and write as follows for 120mm (i.e. B773593-120-X39T).
B773593
(Part No.)
120
("A" length)
X39
(Cable length
in inches)
Connector
(if any)