PPB Sodium ISE TEXT SIZE: A A A    PRINT THIS PAGE   EMAIL THIS PAGE

VLC is pleased to announce our new Sodium Ion Selective Electrode (ISE) capable of measuring down to the PPB (parts per billion) range!

Model # NAB1502
Reproducibility +/- 5%
Interferences Ag+, Li+, H+
Temperature Range 0 to 80 degrees C
Pressure Range 0 to 70 psi
Response Time 95% in 2 minutes
Storage 100 PPB (Parts Per Billion) Sodium standard solution
Concentration Range 0.1 PPB to 10 PPM
pH Range > 11
Resistance < 200 MOhms
Care, Maintenance & Cleaning Rinse with a dilute NH4Cl/NH4OH ISA, followed by soaking in 100 PPB (Parts Per Billion) Sodium standard
Temperature Compensation Yes
Isopotential Point 30 ppm (Parts Per Million) Sodium
On-Line Capability Yes, pH Control required
Applications Power Plants, Boiler Feed Water