Van London – pHoenix Company
6103 Glenmont Dr.
Houston, TX 77081
USA
(Toll Free) 800-522-7920
(voice) 713-772-6641
(fax) 713-772-4671
Request Literature
|
|
|
|
|
|
An Ion Selective Electrode sensor specifically designed to measure Ammonia from 0.01 ppm to 17,000 ppm NH3 (5x10-7 M to 1M) concentrations.
Applications include, but not limited to: High Purity Power Station Water, Fish Tanks, Sea Water, Waste Water, Plating Baths, Biological Samples.
|
|
Model #
|
NH31501C
|
|
Slope
|
56 + / - 3 mV/decade
|
|
Reproducibility
|
+ / - 2%
|
|
Interferences
|
Volatile amines
|
|
Temperature Range
|
0 to 50° C (if calibrated at that temperature)
|
|
Pressure Range
|
0 to 10 psi
|
|
Response Time
|
95% response in 30 seconds
|
|
Storage
|
Long term, store dry Short term, store in ammonia standard with ISA added
|
|
Concentration Range
|
5 x 10-7 M to 1M 0.01 ppm to 17,000 ppm NH3
|
|
pH Range
|
Above 11 pH
|
|
Resistance
|
1000 to 1500 megaohms
|
|
Care, Maintenance & Cleaning
|
Treat with a dilute ammonia standard with pH adjusted > pH 11 Replace membrane as needed
|
|
Temperature Compensation
|
No
|
|
Isopotential Point
|
Unknown
|
|
On-Line Capability
|
Yes, pH control required
|
|
Applications
|
High Purity Power Station Waters, Fish Tanks, Sea Waters, Waste Waters, Plating Baths, Biological Samples
|
|
|