A Cadmium Ion Selective Electrode sensor specifically designed to measure 0.01 ppm to 11,200 ppm (10-7 M to 10-1 M) concentrations of Cd+2.
Applications include, but not limited to: Plating Baths.
|
|
Model #
|
CD21501
|
|
Slope
|
27 + / - 3 mV / decade
|
|
Reproducibility
|
+ / - 2%
|
|
Interferences
|
Ag+, Hg+2, Cu+2, Pb+2, Fe+2
|
|
Complexation
|
Heavy metals
|
|
Temperature Range
|
0 to 80° C (if calibrated at that temperature)
|
|
Pressure Range
|
0 to 70 psi
|
|
Response Time
|
95% response in 30 seconds
|
|
Storage
|
Long term, store dry
Short term, store in dilute cadmium standard
|
|
Concentration Range
|
10-7 to 10-1M
0.01 ppm to 11,200 ppm
|
|
pH Range
|
2 to 12 pH
|
|
Resistance
|
< 1 megaohm
|
|
Care, Maintenance & Cleaning
|
Polish with polishing strips
Treat with a dilute ISA
|
|
Temperature Compensation
|
Not recommended
|
|
Isopotential Point
|
50,000 ppm Cadmium
|
|
On-Line Capability
|
Yes
|
|
Applications
|
Plating Baths
|
|
|