ISE: Copper TEXT SIZE: A A A    PRINT THIS PAGE   EMAIL THIS PAGE

A Copper (Cupric) Ion Selective Electrode sensor specifically designed to measure 0.0006 ppm to 6350 ppm (1x10-8 M to 1x10-1 M) concentrations of Cu+2.
Applications include, but not limited to: Plating Baths, Water.

Model # CU01502
Slope 27 ± 2 mV between 10-100 ppm @ 25°C
Reproducibility ± 2%
Interferences Ag+, Hg+2, Cl-, Br-, Fe+2, Cd+2
Complexation S-2, PO4-3, OH-1
Temperature Range 0 to 80° C (if calibrated at that temperature)
Pressure Range 0 to 70 psi
Response Time 95% response in 30 seconds
Storage Long term, store dry
Short term, store in dilute copper standard
Concentration Range 0.0006 ppm to 6350 ppm  (1x10-8 M to 1x10-1 M)
pH Range 2 to 12 pH
Resistance < 1 Megaohm
Care, Maintenance & Cleaning Polish with polishing strips
Treat with a dilute ISA
Isopotential Point 50,000 ppm
On-Line Capability Yes
Applications Plating Baths, Water