AlpHa Measurement Solutions
10540 Rockley Road
Houston, TX 77099
USA
(Toll Free) 800-522-7920
(voice) 832-456-4100
Request Literature
|
|
Follow us on:
 |
|
|
|
A Fluoride Ion Selective Electrode sensor specifically designed to measure .02 ppm to saturated (1x10-6 M to saturated) concentrations of F-.
Applications include, but not limited to: Drinking Water, Wastewater and Natural Waters, Air and Stack Gases, Acids, Sea Water, Minerals, Soil, Foods, Biological Fluids, Toothpaste.
|
|
Model #
|
F001502
|
Slope
|
56 ± 2 mV between 10-100 ppm @ 25°C
|
Reproducibility
|
± 2%
|
Interferences
|
OH-1
|
Complexation
|
Al+3, Fe+3, H+1
|
Temperature Range
|
0 to 80° C (if calibrated at that temperature)
|
Pressure Range
|
0 to 70 psi
|
Response Time
|
95% response in 30 seconds
|
Storage
|
Long term, store dry
Short term, store in dilute fluoride standard with TISAB added
|
Concentration Range
|
0.02 ppm to Saturated (1 x 10-6 M to Saturated)
|
pH Range
|
5 to 8 pH
|
Resistance
|
100 to 200 kiloohms
|
Care, Maintenance & Cleaning
|
Polish with polishing strips
Treat with a dilute ISA
|
Isopotential Point
|
20 ppm
|
On-Line Capability
|
Yes, pH control required
|
Applications
|