Van London – pHoenix Company
6103 Glenmont Dr.
Houston, TX 77081
USA
(Toll Free) 800-522-7920
(voice) 713-772-6641
(fax) 713-772-4671
Request Literature
|
|
|
An Iodide Ion Selective Electrode sensor specifically designed to measure 0.006 ppm to 127,000 ppm (5x10-8 M to 1M) concentrations of I-.
Applications include, but not limited to: Milk, Feeds andPlants, Pharmaceuticals.
|
|
Model #
|
I001502
|
|
Slope
|
57 + / - 3 mV / decade
|
|
Reproducibility
|
+ / - 2%
|
|
Interferences
|
S-2, Br-, CN-
|
|
Complexation
|
Heavy metals
|
|
Temperature Range
|
0 to 80° C (if calibrated at that temperature)
|
|
Pressure Range
|
0 to 70 psi
|
|
Response Time
|
95% response in 30 seconds
|
|
Storage
|
Long term, store dry
Short term, store in dilute iodide standard
|
|
Concentration Range
|
5 x 10-8 M to 1M
0.006 ppm to 127,000 ppm
|
|
pH Range
|
0 to 14 pH
|
|
Resistance
|
< 1 megaohm
|
|
Care, Maintenance & Cleaning
|
Polish with polishing strips
Treat with a dilute ISA
|
|
Temperature Compensation
|
Not recommended
|
|
Isopotential Point
|
50,000 ppm Iodide
|
|
On-Line Capability
|
Yes
|
|
Applications
|
Milk, Feeds, Plants, Pharmaceuticals
|
|
|