A sensor specifially designed to measure 0.2 ppm to 20,700 ppm (10-6 M to 0.1M) concentrations of Pb+2.
Applications include, but not limited to: Plating Baths.
|
|
Model #
|
PB21502
|
|
Slope
|
26 + / - 3 mV / decade
|
|
Reproducibility
|
+ / - 2%
|
|
Interferences
|
Ag+1, Hg+2, Cu+2, high levels of Cd+2, Fe+2
|
|
Temperature Range
|
0 to 80° C (if calibrated at that temperature)
|
|
Pressure Range
|
0 to 70 psi
|
|
Response Time
|
95% response in 30 seconds
|
|
Storage
|
Long term, store dry
Short term, store in dilute lead standard
|
|
Concentration Range
|
1 x 10-6 M to 0.1M
0.2 ppm to 20,700 ppm
|
|
pH Range
|
3 to 8 pH
|
|
Resistance
|
< 1 Megaohm
|
|
Care, Maintenance & Cleaning
|
Polish with polishing strips
Treat with a dilute ISA
|
|
Temperature Compensation
|
Not recommended
|
|
Isopotential Point
|
50,000 ppm Lead
|
|
On-Line Capability
|
ON-LINE CAPABILITY
|
|
Applications
|
Plating baths
|
|
|