ISE: Magnesium TEXT SIZE: A A A    PRINT THIS PAGE   EMAIL THIS PAGE

Unique to Van London, the Magnesium Ion Selective Electrode sensor is designed to provide greater accuracy in water hardness, drilling muds, mining, drinking water, mineral water, and sea water. The Mg+2 ISE can also be accurately applied to monitoring total water hardness titrations.

Model # MG21502
Slope 25 ± 3 mV between 10-100 ppm at 25°C
Reproducibility ± 2%
Interferences Cu+2, Ni+2, Zn+2, Fe+2
Temperature Range 0 to 40°C (if calibrated at that temperature)
Pressure Range 0 to 30 psi
Response Time 95% response in 30 seconds
Storage Long term, store dry
Short term, store in dilute Magnesium standard
Concentration Range 0.1 ppm to 30,000 ppm
pH Range 3 to 10 pH
Resistance 100 megaohms
Care, Maintenance & Cleaning Treat with distilled water followed by Magnesium standard
Isopotential Point 1 ppm
On-Line Capability Yes
Applications Water Softening Systems, Drinking and Mineral Waters, Sea Water,