ISE: Silver/Sulfide TEXT SIZE: A A A    PRINT THIS PAGE   EMAIL THIS PAGE

A Silver / Sulfide Ion Selective Electrode sensor specifially designed to measure either 0.01 ppm to 107,900 ppm (10-7 M to 1M) concentrations of Silver (Ag+), or 0.003 ppm to 32,100 ppm (10-7 M to 1M) concentrations of Sulfide (S-2).
Applications include, but not limited to: Sewage effluent, Soils, Sediments (Sulfide), Plating Baths (Silver), Photographic Fixing Solution (Silver).

Model # AGS1502
Slope 57 + / - 2 mV / decade (Silver) 27 + / - 2 mV / decade (Sulfide)
Reproducibility + / - 2%
Interferences Hg+1, Hg+2
Complexation Heavy metals
Temperature Range 0 to 80° C (if calibrated at that temperature)
Pressure Range 0 to 70 psi
Response Time 95% response in 30 seconds
Storage Long term, store dry Short term, store in dilute silver(or sulfide) standard
Concentration Range 1 x 10-7 M to 1M 0.01 ppm to 107,900 ppm (Silver) 0.003 ppm to 32,100 ppm (Sulfide)
pH Range 2 to 8 pH (Silver) Above 11 pH (Sulfide)
Resistance < 1 Megaohm
Care, Maintenance & Cleaning Polish with polishing strips Treat with a dilute ISA
Temperature Compensation Not Recommended
Isopotential Point 50,000 ppm Silver, Sulfide is unknown
On-Line Capability Yes, pH control required for sulfide monitoring
Applications Sewage Effluent, Soils, Sediments (Sulfide) Plating Baths, Photographic Fixing Solution (Silver)