Model # |
CLO1502 |
| Slope |
56 + / - 3 mV / decade |
| Reproducibility |
+ / - 2% |
| Interferences |
S-2, I-1, CN-1, Br-1 |
| Temperature Range |
to 80° C (if calibrated at that temperature) |
| Pressure Range |
0 to 70 psi |
| Response Time |
95% response in 30 seconds |
| Storage |
Long term, store dry
Short term, store in dilute chloride standard |
| Concentration Range |
5 x 10-5 M to 1M
1.8 ppm to 35,500 ppm |
| Resistance |
< 1 megaohm |
| Care,Maintenance & Cleaning |
Polish with polishing strips
Treat with a dilute ISA |
| Temperature Compensation |
Not recommended |
| Isopotential Point |
50,000 ppm Chloride |
| On-line Capability |
Yes |
| Applications |
River and Tap Water, Plant Tissue, Soils, Boiler Feed Water, Blood Electrolytes/Clinical Analysis, Sweat, Urine, Blood, Cement, Plating Baths, Food Samples |
|