Model # |
NA71502 |
| Slope |
55 + / - 3 mV / decade |
| Reproducibility |
+ / - 2% |
| Interferences |
Ag+1, Li+1, H+1 |
| Temperature Range |
0 to 80° C (if calibrated at that temperature) |
| Pressure Range |
0 to 70 psi |
| Response Time |
95% response in 30 seconds |
| Storage |
Store in 1 M NaCl |
| Concentration Range |
1 x 10-6 M to 5M
0.02 ppm to 100,000 ppm |
| Resistance |
300 to 500 megaohms |
| Care,Maintenance & Cleaning |
Rinse with a dilute NH4Cl/NH4OH ISA, followed by Soak with 5 M NaCl |
| Temperature Compensation |
Yes |
| Isopotential Point |
30 ppm Sodium |
| On-line Capability |
Yes, pH control required |
| Applications |
Steam Condensates in Power Plants, Blood Electrolytes/Clinical Analysis, Serum, Foods, Wine, Sea Water |
|