2-electrode conductivity sensor with integrated temperature probe (NTC 30 kohm), stainless steel body incl. flow cell. For measurement in low conductivity solutions such as ultrapure water or boiler feed water, e.g. for checking water desalination plants.
Suitable for use with the following portable meters:
Portamess® 913 (X) Cond
Portamess® 911 (X) Cond
|
| Specifications |
| Number of electrodes: |
2 |
Body:
Electrode material: |
Stainless steel 1.4571
Stainless steel 1.4571 |
| Body length: |
120 mm |
| Body diameter: |
12 mm |
| Temperature probe: |
NTC (30 kohm): –5 ... +100 ºC |
| Immersion depth: |
Min.: 30 mm, max.: Total length incl. cable |
| Pressure resistance: |
2 bars |
| Cell constant: |
0.100 cm-1 ±2 % |
| Ranges: |
0,01 ... 200 µS/cm |
|