| Specifications |
| Permissible process pressure and temperature during movement |
| Process adaptation 1.4571/Hastelloy |
10 bars (at 0 … 140 °C) |
| Process adaptation PEEK HD |
10 bars (at 0 … 140 °C) |
| Process adaptation PVDF HD |
10 bars (0 … 120 °C), 6 bars (140 °C) 30 min |
| Process adaptation PEEK/PVDF |
6 bars (0 … 40 °C) falling linearly to 2 bars (120 °C) |
| Process adaptation PP |
6 bars (5 … 30 °C) falling linearly to 1 bar (80 °C) |
| Permissible process pressure and temperature, statically in service position |
16 bars (at 0 … 40 °C), PP 10 bars (at 5 … 20 °C) |
| Ambient temperature |
–10 … +70 °C |
| Ingress protection |
IP 66 |
| Housing material |
stainless steel A2/PP |
| Permissible pressure for probe control |
4 … 7 bars |
| Quality of compressed air |
| Standard |
gemäß ISO 8573-1:2001 |
| Quality class |
3.3.3 oder 3.4.3 |
| Solid contaminants |
class 3 (max. 5 μm, max. 5 mg/m3) |
| Water content for temperatures > 15 °C |
class 4, pressure dew point 3 °C or below |
| Water content for temperatures 5 … 15 °C |
class 3, pressure dew point –20 °C or below |
| Oil content |
class 3 (max. 1 mg/m3) |
| Sensors |
| With gel electrolyte |
Ø 12 mm, length 225 mm with temp detector, Pg 13.5 thread |
| With liquid electrolyte |
Ø 12 mm, length 250 mm with temp detector |
| Process adaptations |
| Flanges, DIN EN 1092-1 |
DN 32 … 100 |
| Flanges, ANSI B 16.5 |
1 1/2" to 3" |
| Dairy pipe, DIN 11851 |
DN 50 … 100 |
| Ingold socket, 25 mm |
25 mm |
| G 2 1/4" |
for Knick ARF 210/30X |
| Connections |
| Outlet |
G 1/8" internal, 3 m EPDM hose |
| For pressurized sensors |
hose connection NW 6, pressure in sensor chamber
0.5 … 1 bar above process pressure (max. 7 bars) |
| For compressed air (control air for retractable fitting) |
push-in connector DN 4/6 |
Immersion depths/
Installation dimensions |
see dimension drawings |
| Process-wetted materials |
see order code |