SensoGate® WA 131 TEXT SIZE: A A A    PRINT THIS PAGE   EMAIL THIS PAGE

Pneumatic retractable fitting for simple commercially available control units or process applications.

Based on the proven SensoGate® modular design principle, the WA 131 retractable fitting is ideal for applications which do not require completely automated control with central media routing but which still want to make use of superior SensoGate® technology.
SensoGate® WA 131 is ideal for automatic or semi-automatic connections to commercially available or customer developed control units.

SensoGate® WA 131 

Pneumatic retractable fitting for simple commercially available control units or process applications

Variable and reasonably priced
Based on the proven SensoGate® modular design principle, the WA 131 retractable fitting is ideal for applications which do not require completely automated control with central media routing but which still want to make use of superior SensoGate® technology.
SensoGate® WA 131 is ideal for automatic or semi-automatic connections to commercially available or customer developed control units.

Facts
– compact, smooth, and dirt-resistant design
– minimum maintenance costs and maximum flexibility thanks to completely modular design
– very simple replacement of gaskets due to divided calibration/cleaning chamber
– increased service life and reduced seal wear
– ATEX approval II 1GD c II for the complete fitting
– cost reduction by simple installation, operation, and maintenance
– very high availability
– protective rinsing of the seals for a long service life
– SensoLock for high safety of operation
– integrated check valves and limit switches
– fast and uncomplicated replacement of calibration chamber and immersion tube
– all maintenance can be performed on site
– cyclone rinsing for optimum cleaning effect
– superior sensor immersion depth
– standard sensor length (225 mm) even for large immersion depths
– multiple process connections available
– flange version with scraper ring
– modular design allows specific replacement of the process-wetted materials: PEEK, PVDF, PP, 1.4571, Hastelloy
– plastic versions with high pressure and temperature resistance due to carbonfiber-reinforced plastics
– special version for sensors with pressurizable liquid electrolyte

  

Specifications
Permissible process pressure and temperature during movement
Process adaptation 1.4571/Hastelloy 10 bars (at 0 … 140 °C)
Process adaptation PEEK HD 10 bars (at 0 … 140 °C)
Process adaptation PVDF HD 10 bars (0 … 120 °C), 6 bars (140 °C) 30 min
Process adaptation PEEK/PVDF 6 bars (0 … 40 °C) falling linearly to 2 bars (120 °C)
Process adaptation PP 6 bars (5 … 30 °C) falling linearly to 1 bar (80 °C)
Permissible process pressure and temperature, statically in service position 16 bars (at 0 … 40 °C), PP 10 bars (at 5 … 20 °C)
Ambient temperature –10 … +70 °C
Ingress protection IP 66
Housing material stainless steel A2/PP
Permissible pressure for probe control 4 … 7 bars
Quality of compressed air
Standard gemäß ISO 8573-1:2001
Quality class 3.3.3 oder 3.4.3
Solid contaminants class 3 (max. 5 μm, max. 5 mg/m3)
Water content for temperatures > 15 °C class 4, pressure dew point 3 °C or below
Water content for temperatures 5 … 15 °C class 3, pressure dew point –20 °C or below
Oil content class 3 (max. 1 mg/m3)
Sensors
With gel electrolyte Ø 12 mm, length 225 mm with temp detector, Pg 13.5 thread
With liquid electrolyte Ø 12 mm, length 250 mm with temp detector
Process adaptations
Flanges, DIN EN 1092-1 DN 32 … 100
Flanges, ANSI B 16.5 1 1/2" to 3"
Dairy pipe, DIN 11851 DN 50 … 100
Ingold socket, 25 mm 25 mm
G 2 1/4" for Knick ARF 210/30X
Connections
Outlet G 1/8" internal, 3 m EPDM hose
For pressurized sensors hose connection NW 6, pressure in sensor chamber
0.5 … 1 bar above process pressure (max. 7 bars)
For compressed air (control air for retractable fitting) push-in connector DN 4/6
Immersion depths/
Installation dimensions
see dimension drawings
Process-wetted materials see order code