Van London-pHoenix Company LogoVan London-pHoenix Company, your source for pH, ORP (Oxidation-Reduction Potential), Conductivity, Oxygen, and Ion Selective Electrodes, for Laboratory, Biotechnology and Medical Applications
pHORPRedoxIon SelectiveConductivityDissolved OxygenOther ProductsTechnical InfoAbout Van London-pHoenix CompanyContact Van London-pHoenix Company

London-pHoenix Co. Customer Login
Username

Password

Not Registered Yet?

 

Van London – pHoenix Company
6103 Glenmont Dr.
Houston, TX 77081
USA

(Toll Free) 800-522-7920
(voice) 713-772-6641
(fax) 713-772-4671

Request Literature

SensoGate® WA 131M TEXT SIZE: A A A    PRINT THIS PAGE   EMAIL THIS PAGE

Innovative manual retractable fitting with safe-to-operate rotary drive.

For applications in which pneumatic fittings are not essential, Knick has developed the SensoGate® WA 131M, a state-of-the-art manual sensor lock-gate fully compatible with its pneumatically controlled “ sister fittings”. This applies to rinsing chambers and process connections as well as the materials used.

SensoGate® WA 131M

Innovative manual retractable fitting with safe-to-operate rotary drive

For applications in which pneumatic fittings are not essential, Knick has developed the SensoGate® WA 131M, a state-of-the-art manual sensor lock-gate fully compatible with its pneumatically controlled “ sister fittings”. This applies to rinsing chambers and process connections as well as the materials used.

Revolutionary operation
The SensoGate® WA 131 M has a manual rotary drive, which enables the fitting to be operated up to a counter pressure of 10 bars – and this in less than two turns. Regardless of how unfavorable the angle or the operating height at which the fitting is installed, the low rotary forces can easily be applied using the wrist. This is not possible with the sliding motion of conventional fittings.

Maximum process safety
The configuration of the O-rings means that there is never any connection between the drain opening of the rinsing chamber and the process, even during the inward and outward movement. For reasons of safety, the sensor can only be removed in the service position. Automatic extension protection reliably blocks the movement of the fitting and prevents extension into the process position with no sensor. Furthermore, the immersion tube is secured against twisting, which guarantees constant flow conditions for the sensor in the process.

Facts
– compact, smooth, and dirt-resistant design
– minimum maintenance costs and maximum flexibility thanks to completely modular design
– very simple replacement of seals due to divided calibration/cleaning chamber
– increased service life and reduced seal wear
– ATEX approval II 1GD c II for the complete fitting
– cost reduction by simple installation, operation, and maintenance
– very high availability
– protective rinsing of the seals for a long service life
– SensoLock for high safety of operation
– integrated limit switches
– fast and uncomplicated replacement of calibration chamber and immersion tube
– all maintenance can be performed on site
– cyclone rinsing for optimum cleaning effect
– superior sensor immersion depth
– standard sensor length (225 mm) even for large immersion depths
– multiple process connections available
– flange version with scraper ring
– modular design allows specific replacement of the process-wetted materials: PEEK, PVDF, PP, 1.4571, Hastelloy
– plastic versions with high pressure and temperature resistance due to carbonfiber-reinforced plastics
– special version for sensors with pressurizable liquid electrolyte

 

Specifications
Permissible process pressure and temperature during movement
Process adaptation 1.4571/Hastelloy 10 bars (at 0 … 140 °C)
Process adaptation PEEK HD 10 bars (at 0 … 140 °C)
Process adaptation PVDF HD 10 bars (0 … 120 °C), 6 bars (140 °C) 30 min
Process adaptation PEEK/PVDF 6 bars (0 … 40 °C), falling linearly to 2 bars (120 °C)
Process adaptation PP 6 bars (5 … 30 °C), falling linearly to 1 bar (80 °C)
Permissible process pressure and temperature, statically in service position 16 bars (at 0 … 40 °C), PP 10 bars (at 5 … 20 °C)
Ambient temperature –10 … +70 °C
Ingress protection IP 66
Housing material Stainless steel A2/PP or stainless steel/PEEK
Permissible pressure for probe control 4 … 7 bars
Sensors
With gel electrolyte Ø 12 mm, length 225 mm with temp detector, Pg 13.5 thread
With liquid electrolyte Ø 12 mm, length 250 mm with temp detector
Process adaptations
Flanges, DIN EN 1092-1 DN 32 … 100
Flanges, ANSI B 16.5 1 1/2" to 3"
Dairy pipe, DIN 11851 DN 50 … 100
Ingold socket, 25 mm 25 mm
G 2 1/4" for Knick ARF210/30X
Connections
Rinse and calibration media (optional) internal thread G 1/8", hose coupling G 1/8"-DN 4/6
Outlet (optional) outlet hose, NW 8 EPDM 3m
For pressurized sensors hose connection NW 6, pressure in sensor chamber
0.5 …1 bar above process pressure (max. 7 bars)
Immersion depths/
Installation dimensions
see dimension drawings, page 261
Process-wetted materials see order code, page 258