SZ 7251 Autoclean Flow Cell TEXT SIZE: A A A    PRINT THIS PAGE   EMAIL THIS PAGE

Auto cleaning flow cell designed for the in-line or in-flow continuous measurement of Residual Chlorine or Dissolved Ozone in solution.

SZ 7251 Auto clean flow cell for Free Chlorine / Dissolved Ozone

Specifications

Transparent body: acrylic
Holder: PVC
O Ring: NBR and fluoridated elastomer
Fittings: polypropylene
Tubing: polythene
Balls: N° 20 into the body
Inlet/outlet: 1/8 " fittings
Diameter: 40 mm max.
Length: 150 mm max.
Flow: 15/40 liter/hour constant
Temperature: 0/50 °C
Connecting tubing: Diameter 4X6 mm. L=5 m
Sensor: to be ordered separately depending on the application

This cell is designed for the in-line or in-flow continuous measurement of Residual Chlorine or Dissolved Ozone in solution.

The measuring sensor is inserted in the holder of the cell, which protects the body and places the sensing part in the right position into the cell.

The sample inlet flow will create a continuous movement of the internal balls whose contact with the sensing part of the sensor will perform a self cleaning action.

The package includes:
- plastic tube for the connection to the sample - fixing clamp - spare balls

The sample inlet pressure must be constant in order to get a continuous flow necessary to obtain stable and reliable measuring.