Van London Company LogoVan London Company, your source for pH, ORP (Oxidation-Reduction Potential), Conductivity, Oxygen, and Ion Selective Electrodes, for Laboratory, Biotechnology and Medical Applications
pHORPRedoxIon SelectiveConductivityDissolved OxygenOther ProductsTechnical InfoAbout Van London-pHoenix CompanyContact Van London-pHoenix Company

London-pHoenix Co. Customer Login
Username

Password

Not Registered Yet?

 

AlpHa Measurement Solutions
10540 Rockley Road
Houston, TX 77099
USA

(Toll Free) 800-522-7920
(voice) 832-456-4100

Request Literature

 
Follow us on:
   
SZ 283 Potentiostatic electrode TEXT SIZE: A A A    PRINT THIS PAGE   EMAIL THIS PAGE

This sensor is designed for the measurement of Free Chlorine, Chlorine dioxide and Dissolved Ozone in water.

Note: This specific model of potentiostatic chlorine sensor is currently not available to OEM customers, and is only available when ordered as a complete package with the B&C CL 7685 Potentiostatic Cl2 Controller and SZ 7231 - SZ 7233 Flow Cells.

SZ 283
Potentiostatic electrode

SZ 283 Potentiostatic electrode

This sensor is designed for the measurement of Free Chlorine, Chlorine dioxide and Dissolved Ozone in water.

The potentiostatic method is an "amperometric" measurement with constant potential, made through 2 metal electrodes and a reference electrode dipped in a cell.

The current running through the cell consumes Chlorine or Ozone contents, thus requiring constant renewal through a constant liquid flow.

In the traditional amperometric measurement it remains difficult to maintain a constant relation between cell current and Chlorine / Ozone concentration, especially near zero, because of the ORP and liquid resistance effects.  As a result, frequent zero and sensitivity calibration are needed.

In the potentiostatic measuring, the electrodes' potential is electronically controlled in relation to the liquid, providing a linear current/concentration relationship and a very stable zero value in oxidative absence.

The sensor is shaped so that it is easy to clean and replace.

It is recommended to place the sensor in measurement cell SZ 7231 or SZ 7233 provided with overflow in order to maintain constant sample flow.

The recommended meter(s) for this sensor are the CL 7635 or CL 7335 controllers.

 

Specifications

Electrodes: 2 Platinum rings
Reference: gel with annular junction
Body: glass
Cable: 3 m
Max pressure: 10 bar at 20°C
Dimensions: 110x12 mm