READINGS ARE NOT REPRODUCIBLE |
- Is there sample carryover?
- Are there sample interferences or complexing agents present?
- Is the reference electrode junction contaminated?
|
SLOW RESPONSE (READINGS SLOWLY CHANGING) |
- Was the electrode stored in the wrong solution?
- Was the electrode poisoned by the sample?
|
OUT OF RANGE READING |
- Is the electrode plugged into the controller properly?
- Is there a reference electrode present?
- Is there enough fill solution left in the reference electrode?
- Is the electrode placed in the sample solution?
|
LOW SLOPE OR NO SLOPE |
- Are the standards too old, contaminated, or made wrong?
- Has the sample pH been adjusted properly to the operating range of the electrode?
- Is there an air bubble on electrode surface?
- Is the controller okay?
- Is there enough fill solution left in the reference electrode?
|
NOISY RESPONSE (READINGS RANDOMLY CHANGING) |
- Is the controller grounded?
- Is there an air bubble on electrode surface?
- Is the controller operating properly?
- Is there enough fill solution left in the reference electrode?
|
DRIFTY RESPONSE (READINGS CONTINUOUSLY CHANGING) |
- Is there excessive leaking at the reference electrode junction?
- Is the reference electrode junction clogged?
- Was the sensing membrane poisoned by the sample?
- Are there temperature problems?
- Is the sample too concentrated?
- Does the sensing membrane need conditioning?
|
INACCURATE READINGS (BUT CALIBRATION IS OK) |
- Are the standards incorrect?
- Was the sample pH adjusted properly?
- Is there sample carryover?
- Are there sample interferences or complexing agents present?
|