Van London-pHoenix Company LogoVan London-pHoenix Company, your source for pH, ORP (Oxidation-Reduction Potential), Conductivity, Oxygen, and Ion Selective Electrodes, for Laboratory, Biotechnology and Medical Applications
pHORPRedoxIon SelectiveConductivityDissolved OxygenOther ProductsTechnical InfoAbout Van London-pHoenix CompanyContact Van London-pHoenix Company

London-pHoenix Co. Customer Login
Username

Password

Not Registered Yet?

 

Van London – pHoenix Company
6103 Glenmont Dr.
Houston, TX 77081
USA

(Toll Free) 800-522-7920
(voice) 713-772-6641
(fax) 713-772-4671

Request Literature

ISE: Chloride TEXT SIZE: A A A    PRINT THIS PAGE   EMAIL THIS PAGE

A Chloride Ion Selective Electrode sensor specifically designed to measure 1.8 ppm to 35,500 ppm (5x10-5 M to 1M) concentrations of Cl-.
Appliations include, but not limited to: River and Tap Water, Plant Tisue, Soils, Boiler Feed Water, Blood Electrolytes/Clinical Analysis, Sweat, Urine, Cement, Plating Baths, Food Samples.

Model # CLO1502
Slope 56 + / - 3 mV / decade
Reproducibility + / - 2%
Interferences S-2, I-1, CN-1, Br-1
Complexation Heavy metals
Temperature Range to 80° C (if calibrated at that temperature)
Pressure Range 0 to 70 psi
Response Time 95% response in 30 seconds
Storage Long term, store dry Short term, store in dilute chloride standard
Concentration Range 5 x 10-5 M to 1M 1.8 ppm to 35,500 ppm
pH Range 2 to 12 pH
Resistance < 1 megaohm
Care, Maintenance & Cleaning Polish with polishing strips Treat with a dilute ISA
Temperature Compensation Not recommended
Isopotential Point 50,000 ppm Chloride
On-Line Capability Yes
Applications River and Tap Water, Plant Tissue, Soils, Boiler Feed Water, Blood Electrolytes/Clinical Analysis, Sweat, Urine, Blood, Cement, Plating Baths, Food Samples